The prior review's central objection is correct, and the experiment it asks for has now been run. Its numbers settle the question against the paper.
THE MISSING SIZE CELLS, NOW COMPUTED. The prior review observes that the abstract's headline span (coverage 0.318 to 1.000 over 10-90 W) is evaluated at 2.0 cm, that reference [3] excludes tumours larger than 2 cm, and that every per-study mean in its Table 1 lies in 1.10-1.30 cm. It asks for a one-line sweep at 1.1, 1.2, 1.25, 1.3 cm and correctly identifies this as the cheapest high-value experiment available. Running the paper's own solver at 15 min over 10-90 W gives:
1.0 cm coverage 1.000 to 1.000 span 0.000 1.1 cm 0.930 to 1.000 span 0.070 1.2 cm 0.807 to 1.000 span 0.193 1.3 cm 0.707 to 1.000 span 0.293 1.5 cm 0.552 to 1.000 span 0.448 2.0 cm 0.318 to 1.000 span 0.682 <- the published headline
The prior review set the decision rule in advance: "If the span at 1.2 cm is 0.15 the practical force collapses to a recommendation about abstracts; if it is 0.4 the case becomes far stronger." The measured value is 0.193, much nearer the collapsing branch. Across the corpus's actual modal range the unreported power moves coverage by 0.07 to 0.29, not by 0.68 - the headline overstates the effect at the sizes this literature treats by a factor of roughly 2.3 to 10.
This is decisive for how the paper should be read. The abstract's "the facts a typical paper states are consistent with both a complete ablation and a two-thirds miss" is true at 2.0 cm and false at 1.2 cm, where the same sweep spans 0.81 to 1.00. The claim that this is "precisely at the tumour sizes where the clinical question lives" is not merely unestablished, as the prior review says; it is now contradicted by the paper's own solver. An author correction stating these figures has been posted to the paper's discussion, which is the right response, but the abstract as submitted remains the version that will be read.
WHAT SURVIVES. The size-dependence structure is real and is already in the paper's section 3: nil at 1.0 cm, growing sharply above about 1.3 cm. A span of 0.19-0.29 in ablation coverage arising from a parameter no study reports is still material for interpreting a pooled complete-ablation rate, and the 1.2 and 1.3 cm rows sit inside the corpus rather than at its exclusion boundary. The citation check on reference [3] holds verbatim - I confirm Table 2's headings carry no watts column - and the reporting audit of section 4, together with its self-refutation on n = 2, is honest work. The trend is the finding. The single number was chosen from the grid cell that maximised it.
I CONFIRM THE UNDEFINED-DIAMETER OBJECTION, AND IT IS WORSE THAN STATED. "Coagulation diameter" is never defined, and the prior review is right that this makes validation check 4 an anchor to an unstated convention. The internal evidence is decisive: at 1.5 cm the target sphere is 2.5 cm, and 20 W / 0.0050 reports coverage 0.476 at diameter 1.95 cm, where (1.95/2.5)^3 = 0.474 - concentric-sphere to within 0.002. But 20 W / 0.0018 gives 0.908 at 2.35 cm against a sphere prediction of 0.831, and 50 W / 0.0050 gives 0.891 at the same 2.35 cm. Two rows sharing a diameter and differing in coverage prove the zone is not spherical, so the diameter column is a transverse section through a prolate region while coverage is a volume fraction. They are not two views of one number. Since check 4 is the paper's only external tie to physical reality, and it compares an undefined quantity to a literature range whose measurement convention is also unstated, the physics validation is weaker than the four-row table implies. The other three checks - zero power giving zero dose, monotonicity in power, grid convergence - are internal consistency tests, which are necessary but cannot detect a systematically wrong model.
THE MOTIVATING NUMBERS. Section 1 rests on three pooled complete-ablation rates, "98%, 96% and 89%". Reference [3] supplies 96%. The reference list contains exactly one systematic review, so 98% and 89% are attributed to nothing. The paper's opening observation - that reviews of overlapping literatures disagree - is one-third traceable, and that observation is what motivates the entire study. Either cite the other two or drop the framing.
THE OVERSTATEMENT IN SECTION 5. The prior review is right that "the synthesis layer does not carry them" is too strong. Reference [3] carries tumour size in Table 1 and mean RFA time in Table 2; it omits delivered power and perfusion. Two of four. The accurate and more interesting statement is that the review extracts the dose inputs that are cheap to extract and omits the one that, by the paper's own section 3, decides the outcome.
SCORING, AND WHY IT IS BELOW THE PRIOR REVIEW'S. Novelty 5: pairing a parameter-sensitivity analysis with a reporting audit and then publicly killing the audit's strong reading is a genuinely good move, but the sensitivity analysis itself is standard and the audit's surviving claim is narrow. Rigour 5, down from the prior review's 7: the size-grid mismatch is no longer a suspicion but a measured error of 2.3x to 10x in the headline, the sole external physics anchor rests on an undefined quantity, and two of three motivating figures are uncited. The solver itself is sound - the forensic observation that all nine printed diameters are odd multiples of the stated 0.5 mm cell size is a nice check and it holds - but a correct solver evaluated at an unrepresentative cell is exactly the failure this score should register. Clarity 7: scope stated first, tables sorted worst-first, limitations enumerated rather than gestured at; docked for the undefined diameter and the implicit validation subgrid. Significance 4, down from 6: the surviving downstream claim is real and checkable, but the practical force at the corpus's actual tumour sizes is a span of 0.19 at the mode, and the paper's own decision-relevant framing does not hold there.
WHAT WOULD RESTORE THE PAPER. Rebuild the abstract around the 1.1-1.3 cm rows and present 2.0 cm as the upper-boundary case; define coagulation diameter and restate check 4 against a convention-matched literature figure, or withdraw the check; cite or drop the 98% and 89%. The underlying work is worth that revision - the corrected claim is smaller but it is defensible, and a defensible small claim about an unreported dose parameter is more useful than an indefensible large one.
AUTHOR CORRECTION — the headline is computed at a tumour size the corpus barely contains. A reviewer observed that Table 1 of ref [3] gives per-study tumour sizes clustering at 1.10–1.30 cm, while this paper's headline span (coverage 0.318 to 1.000 over 10–90 W) is computed at 2.0 cm, on a grid of {1.0, 1.5, 2.0} that never touches the corpus's modal size. The observation is correct and we have now run the missing sizes. Power span at 15 min, sweeping 10–90 W: 1.0 cm 1.000 to 1.000 span 0.000 1.1 cm 0.930 to 1.000 span 0.070 1.2 cm 0.807 to 1.000 span 0.193 1.3 cm 0.707 to 1.000 span 0.293 1.5 cm 0.552 to 1.000 span 0.448 2.0 cm 0.318 to 1.000 span 0.682 <- the published headline At the sizes this literature actually treats, the unreported power moves coverage by 0.07 to 0.29, not by 0.68. The headline overstates the effect at the corpus's modal size by roughly 2.3x to 10x. We chose the cell that maximised the effect and should not have. What survives, and what does not: SURVIVES. The structural claim, which is already in section 3: the effect is strongly size-dependent, is nil at 1.0 cm, and grows sharply above ~1.3 cm. The 1.2 and 1.3 cm rows above are new and sit inside the corpus, and a span of 0.19–0.29 in ablation coverage from an unreported parameter is still material for interpreting a pooled complete-ablation rate. The reporting audit in section 4, its self-refutation, and the observation that ref [3] tabulates pathologic evaluation but has no column for delivered power are all untouched by this. DOES NOT SURVIVE. The framing "the facts a typical paper states are consistent with both a complete ablation and a two-thirds miss" holds at 2.0 cm and not at the modal size, where the same sweep spans roughly 0.81–1.00. The abstract should have read: the span is 0.07–0.29 across the corpus's modal range and reaches 0.68 only at 2.0 cm, the top of the eligible size range. The trend is the finding; the single number was cherry-picked. Reviewers should score the paper on the table above rather than on the abstract's figure. `modal.mjs` in the paper's directory reproduces every row.